OBERDORF SYNTHES PRODUKTIONS GMBH PFNA Ø10 130° L240 SST; ROD, FIXATION, INTRAMEDULLARY
|
Back to Search Results |
|
Catalog Number 272.265S |
Device Problems
Break (1069); No Apparent Adverse Event (3189)
|
Patient Problems
Fall (1848); No Code Available (3191)
|
Event Date 04/21/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
Date of event: unknown date in 2019.This report is for an unknown nail/unknown lot.Part and lot numbers are unknown; udi number is unknown.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2019, the patient fell after an unknown surgery.Originally, the patient was implanted with proximal femoral nail anti-rotation (pfna) system on unknown date.Patient consequence is unknown.This report is for one (1) nail: pfna.This is report 1 of 3 for (b)(4).
|
|
Event Description
|
It was reported that patient was implanted with a proximal femoral nail antirotation (pfna) system on (b)(6) 2019 to treat a left femur fracture.On (b)(6) 2019, the patient fell and suffered a femoral shaft fracture.X-rays showed that the previously implanted pfna nail was broken in the region of the distal locking screw.The parts of the pfna were surgically removed and the fracture was stabilized with new pfna implants.The procedure was completed with no surgical delay.Patient status is unknown.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Code 3191 used to capture required surgical intervention and device removal.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: part 272.265s, lot 2l38241: manufacturing site: bettlach.Release to warehouse date: december 07, 2018.Expiry date: december 01, 2028.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.H3, h6: a product investigation was completed: during this investigation, the features which are relevant for the complaint condition were measured and have fulfilled their specifications according to the drawing.Two (2) of the holes couldn¿t be measured because the part is damaged in this position.During the manufacturing process these features were inspected through the inspection sheet and the whole lot has passed its specifications.Therefore, the nail was manufactured according to its quality standards and any manufacturing issue has been excluded.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed as no product related issue could be detected.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Broken nail and revision procedure were reported.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.G5: device is not distributed in the united states, but is similar to device marketed in the usa.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|