Model Number 503-50-022
Health Effect - Clinical Code No Code Available (3191)
Date of Event 04/29/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was due to loosening.The previous surgery and the revision detailed in this investigation occurred over 10 years and 9 months apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.Initial or prolonged hospitalization was required.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to loosening.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the patient's event.There are many factors that may contribute to the event that are outside the control of djo surgical such as: loose joints from inadequate soft tissue support, degenerative bone, patient bone deterioration, patient activities or trauma.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Revision surgery - due to a loose glenoid.The surgeon needed to convert to a reverse.
 
Additional Manufacturer Narrative
Corrected data: see brand name, common device name, concomitant medical products, and device manufacture date.Manufacturer narrative: the reason for this revision surgery was reported as loosening.The previous surgery and the revision detailed in this investigation occurred over 10 years and 9 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.Initial or prolonged hospitalization was required.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported components used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the event.The device and its applicable concomitant device was within its respective expiration date at the time of use during the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to loosening.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the patient's event.There are many factors that may contribute to the event that are outside the control of djo surgical are loose joints from inadequate soft tissue support, degenerative bone, patient bone deterioration, patient activities or trauma.No additional information was submitted with the complaint regarding pre-existing conditions of the patient or any activities that may have contributed to the event and hence a definitive root cause cannot be determined.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION SHOULDER
Common Device Name
GLENOID, UNIFORM, PEGGED 50MM CM FOUNDATION SHOULDER
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445 TX 78758 6313
MDR Report Key8653725
Report Number1644408-2019-00563
Device Sequence Number2299103
Product Code KWS
UDI-Device Identifier00888912143608
UDI-Public(01)00888912143608
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K960906
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2013
Device Model Number503-50-022
Device Catalogue Number500-04-150
Device Lot Number53926394
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/29/2019
Supplement Date Received by Manufacturer08/02/2019
Initial Report FDA Received Date05/30/2019
Supplement Report FDA Received Date08/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
500-04-150, LOT 53926394; 503-00-008 LOT 53926935; 503-00-008, LOT 53926935; 503-50-022 LOT 53916882; 500-04-150, LOT 53926394; 503-00-008, LOT 53926935
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age67 YR
-
-