• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. VISUALASE COOLED LASER APPLICATOR SYSTEM; LASER INSTRUMENT, SURGICAL, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. VISUALASE COOLED LASER APPLICATOR SYSTEM; LASER INSTRUMENT, SURGICAL, POWERED Back to Search Results
Model Number 9735560
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/17/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient weight not available.No parts have been received by the manufacturer for evaluation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding a thermal therapy system being used for a neuro soft tissue ablation procedure.It was reported that intra-operatively, saline would not pass from the drip chamber into the tubing.A picture was provided noting that the clip was released and it would not drip through with the pump running or after it was removed.The tubing was replaced and the case proceeded.There was a five minute delay to the procedure due to this issue.There was no reported impact to patient outcome.
 
Additional Manufacturer Narrative
The catheter was returned to the manufacturer for analysis.Analysis found that the returned tubing does not appear to be damaged but does not have flow.Analysis found that the reported event was related to a flow issue.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.Patient weight provided.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISUALASE COOLED LASER APPLICATOR SYSTEM
Common Device Name
LASER INSTRUMENT, SURGICAL, POWERED
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key8653886
Report Number1045254-2019-00269
Device Sequence Number2246675
Product Code GEX
UDI-Device Identifier00643169933385
UDI-Public00643169933385
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K053087
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/07/2021
Device Model Number9735560
Device Catalogue Number9735560
Device Lot Number0216891812
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/11/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/17/2019
Supplement Date Received by Manufacturer06/10/2019
Initial Report FDA Received Date05/30/2019
Supplement Report FDA Received Date07/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age5 YR
Patient Weight23
-
-