MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number TRM-295
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 02/01/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This report is submitted beyond 30-calendar days from allergan¿s receipt due to the exemption transition period.Information contained in this report was previously submitted through psr on 22/apr/2019.A review of the device history record has been completed.No deviations or non-conformance's noted.Device evaluation: analysis of the returned device identified weight of the device within specification, fold creases, and a deformation.There was a cloudy color observed in the gel after the autoclave cycle.Based on the device analysis, the final assessment is no issues found related with the manufacturing process.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The reason for reoperation is capsular contracture, baker grade iii.This is a known potential adverse event addressed in the product labeling.
 
Event or Problem Description
Healthcare professional reported left side capsular contracture, baker grade iii and double capsule.Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key8654201
Report Number9617229-2019-04479
Device Sequence Number2286629
Product Code FTR
UDI-Device Identifier10888628008076
UDI-Public10888628008076
Combination Product (Y/N)N
Initial Reporter StateMI
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/23/2020
Device Catalogue NumberTRM-295
Device Lot Number2857580
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/16/2019
Initial Report FDA Received Date05/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age38 YR
-
-