• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG. MONOPOLAR ELECTROSURGICAL PROBE, L-TIP, 33CM; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG. MONOPOLAR ELECTROSURGICAL PROBE, L-TIP, 33CM; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Catalog Number 0250040112
Health Effect - Clinical Code Burn, Thermal (2530)
Date of Event 04/30/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that the patient was burned.
 
Additional Manufacturer Narrative
Alleged failure: it has been reported by the customer that during a laparoscopic cholecystectomy the patient was burned due to microperforations in the protective.Sheath of the product.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause could be third party repair the product was returned for investigation and the failure mode will be monitored for future reoccurrence.The device manufacturer date is not known.
 
Event or Problem Description
It was reported that the patient was burned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG. MONOPOLAR ELECTROSURGICAL PROBE, L-TIP, 33CM
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key8655066
Report Number0002936485-2019-00224
Device Sequence Number1831800
Product Code GCJ
UDI-Device Identifier37613327061346
UDI-Public37613327061346
Combination Product (Y/N)N
PMA/510(K) Number
K934094
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0250040112
Device Lot Number300816
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2019
Initial Date Received by Manufacturer 05/17/2019
Supplement Date Received by Manufacturer05/17/2019
Initial Report FDA Received Date05/30/2019
Supplement Report FDA Received Date08/27/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-