• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY FRANCE SAS 3003895575 BROACH CORAIL AMT 16; HIP INSTRUMENTS : BROACHES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY FRANCE SAS 3003895575 BROACH CORAIL AMT 16; HIP INSTRUMENTS : BROACHES Back to Search Results
Catalog Number L20416
Medical Device Problem Codes Material Twisted/Bent (2981); Scratched Material (3020)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that after our instruments came out of the wash, it was found that the 25mm drill bit was stuck in the drill guide.As well, the first level of a corail tray was damaged in sterile processing.Finally, the post on the size 16 corail broach was bent while sterile processing.Surgery was not delayed.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: examination of the returned instrument confirmed the reported observation and also found some scratching on the instrument.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BROACH CORAIL AMT 16
Common Device Name
HIP INSTRUMENTS : BROACHES
Manufacturer (Section D)
DEPUY FRANCE SAS 3003895575
7 allée irène joliot curie
bp 256
saint priest cedex 69801
FR  69801
MDR Report Key8655159
Report Number1818910-2019-94834
Device Sequence Number2473244
Product Code HTQ
UDI-Device Identifier10603295258186
UDI-Public10603295258186
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberL20416
Device Lot NumberA0311
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/24/2019
Initial Date Received by Manufacturer 05/13/2019
Supplement Date Received by Manufacturer05/24/2019
07/09/2019
Initial Report FDA Received Date05/30/2019
Supplement Report FDA Received Date06/06/2019
07/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-