D10.Device available, return date: additional information: g4.Date manufacturer received; additional information: g7.Type of report; additional information: h2.Follow-up type; additional information: h3.Device evaluation, device returned; additional information: h6.Evaluation codes; additional information: device evaluation; h10.Additional manufacturer narrative; additional information: device evaluation, the device was returned as part of this investigation.Analysis found the concerto coil was returned within an introducer sheath.There was no instant detacher, microcatheter, or accessories returned with the device.The concerto coil was decontaminated.There is no evidence of any detachment attempts by mechanical or manual methods.The concerto pusher was found broken approximately 142cm from the proximal end with the release wire extending out of the distal end.The distal segment of the release wire was broken with the coin, lumen stop, retainer ring, shield coil, detach element and implant coil missing and not returned for evaluation.The investigation determined that there was insufficient information to determine the cause of the event.Since the microcatheter, distal segment of release wire, coin, lumen stop, retainer ring, shield coil, detach element, and implant coil were not returned for analysis, any contribution of these subassemblies to the coil separation/break could not be determined.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|