| Catalog Number |
TEXP130RUH |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
04/22/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) year old female patient who underwent breast reconstruction with a 650cc artoura breast tissue expander on the left side, suffered leaking postoperatively.As a result, the patient underwent removal on (b)(6) 2019.
|
| |
|
Additional Manufacturer Narrative
|
|
On 6/5/2019, the mentor failure analysis lab has received the device for evaluation.On 6/20/2019, the product investigation was completed.Device investigation summary: during evaluation of the sample it appeared intact.Leak testing was performed and it revealed a tear in the patch, measurement approximately less than 0.1 cm.Microscopic examination was performed and no evidence of instrument damage was observed.No other anomalies were observed.The manufacturing records were reviewed, and the manufacturing criteria were met prior to the release of this lot.Deflation is a known complication associated with mentor mammary prostheses.Possible causes of deflation of saline-filled implants include, but are not limited to the following events: excessive stresses or manipulations as may occur during normal, daily routines including customary and purposeful trauma to the breast.Breast implants are not considered lifetime devices and deflation is a known complication associated with these devices and is referenced in our product insert data sheet.It should be noted that as part of our quality process, each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.Based on the information reported and/or the product investigation, there is no evidence that the issue is related with the manufacturing process.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|