| Model Number |
LVZRXT180J |
| Medical Device Problem Code |
Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/15/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product event summary: the guidewire was returned and analyzed.The guidewire was unraveled.The distal conductor of the lead was obstructed due to a guidewire stuck in the lumen.Visual analysis of the guidewire indicated damage during use.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during left ventricular (lv) lead implant the guidewire got stuck after being placed and maneuvered inside the lead, making it hardly removable.The lead was removed and a new lead was implanted.The guidewire was returned, analyzed, and tested out of specification.No patient complications have been reported as a result of this event.
|
| |
|