MEDTRONIC, INC ZINGER XT; OCCLUDER, CATHETER TIP Back to Search Results
Model Number LVZRXT180J
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/15/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product event summary: the guidewire was returned and analyzed.The guidewire was unraveled.The distal conductor of the lead was obstructed due to a guidewire stuck in the lumen.Visual analysis of the guidewire indicated damage during use.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during left ventricular (lv) lead implant the guidewire got stuck after being placed and maneuvered inside the lead, making it hardly removable.The lead was removed and a new lead was implanted.The guidewire was returned, analyzed, and tested out of specification.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZINGER XT
Common Device Name
OCCLUDER, CATHETER TIP
Manufacturer (Section D)
MEDTRONIC, INC
37a cherry hill dr
danvers MA 01923
MDR Report Key8656042
Report Number1220452-2019-00071
Device Sequence Number2473172
Product Code DQT
UDI-Device Identifier00643169452299
UDI-Public00643169452299
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K983927
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/24/2021
Device Model NumberLVZRXT180J
Device Catalogue NumberLVZRXT180J
Device Lot NumberG18A10971
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/30/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/17/2019
Initial Report FDA Received Date05/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age83 YR
-
-