MAUDE Adverse Event Report: BARD ACCESS SYSTEMS SITE~RITE 8 ULTRASOUND SYSTEM CONSOLE WITH CONNECTOR; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS SITE~RITE 8 ULTRASOUND SYSTEM CONSOLE WITH CONNECTOR; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770600
Device Problem Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The manufacturer has received the sample and will be evaluated.Supplemental will be submitted with evaluation results.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
The visual on the screen is getting worse as time goes on.
 
Manufacturer Narrative
H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.The sr8 was tested with an in house gt probe and showed no signs of lagging behind with any movement and the image looked normal compared to other sr8 units.During evaluation, the reported issue of poor ultrasound image is inconclusive as the reported issue could not be reproduced during evaluation.No other functionality issues with the equipment were found during evaluation/servicing.The device was serviced, tested, and returned to the customer.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
The visual on the screen is getting worse as time goes on.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE~RITE 8 ULTRASOUND SYSTEM CONSOLE WITH CONNECTOR
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8656734
MDR Text Key146633338
Report Number3006260740-2019-01519
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741138263
UDI-Public(01)00801741138263
Combination Product (y/n)N
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 08/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770600
Device Catalogue Number9770600
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/14/2019
Event Location Hospital
Initial Date Manufacturer Received 05/09/2019
Initial Date FDA Received05/30/2019
Supplement Dates Manufacturer Received07/15/2019
Supplement Dates FDA Received08/09/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-