|
All siemens ahbs2 controls following ahbs2 reagent lots 104 and 106 were within normal ranges.No instrument errors were found with the system during events.The cause for the discordant ahbs2 results is unknown.Siemens healthcare diagnostics is investigating.Mdr 1219913-2019-00100 (initial result), mdr 1219913-2019-00101 (repeat 1 result), and mdr 1219913-2019-000102 (repeat 2 result) were filed for the same event.
|
|
Siemens filed the initial mdr (b)(4) on (b)(6), 2019.(b)(6) 2019 additional information: the customer had repeated the sample on reagent lot # 106 because there was not enough reagent lot # 104 left.The type of blood collection tube used was plastic gel barrier tubes made in china.The patient serum sample cannot be returned to siemens for further evaluation due to china customs issues.Siemens is awaiting further information from the customer.Mdr (b)(4) supplemental report 1 (initial result), mdr (b)(4) supplemental report 1 (repeat 1 result), and mdr (b)(4) supplemental report 1 (repeat 2 result) were filed for the same event.
|
|
Siemens filed the initial mdr(b)(4) on (b)(6)2019.Siemens filed the mdr (b)(4) supplemental report 1 on (b)(6)2019.06/20/2019 additional information: the customer had a sample that recovered reactive (246.58 miu/ml) with advia centaur xp anti-hbs2 (ahbs2) lot 104 but when repeated with advia centaur xp ahbs2 lot 106 the sample recovered nonreactive (<3.1 and <3.1 miu/ml).When the sample was retested with lot 104 it recovered repeatedly reactive (258.42, 263.6 and 257.4 miu/ml).Many samples were measured with ahbs2 lot 104 and remeasured with lot 106 on that day, similar issues were not found with these samples.No instrument errors were found with the system during events.The sample cannot be sent to siemens for further investigation due to chinese regulations.The customer is not able to provide the patient's medical status other than gastro-intestinal bleeding or a list of medications/supplements the patient is taking.The customer considers the reactive result to be correct but the sample has not been tested with another anti-hbs method to verify which result is correct.The sample was also hbsii reactive (>1000 index), hbct reactive (>8 index), hbeag nonreactive (0.01 index), and ahbe reactive (>4.5 index).If the patient had a resolved infection, typically we would expect ahbs2 to be reactive, ahbe to be reactive, and hbct to be reactive but we would not typically expect the hbsii result to be >1000 index.If the patient had a late acute infection, typically we would expect hbsii to be reactive, ahbe to be reactive, and hbct to be reactive but we would not typically expect the ahbs2 result to be >200 miu/ml.The clinical sensitivity and specificity section of the advia centaur xp anti-hbs2 instructions for use (ifu) (10629819 revision k) lists the 95% confidence interval (ci) for resolved relative sensitivity as 98.30% - 100%, so a certain number of false negatives can be expected with a given lot.If the advia centaur xp ahbs2 lot 106 result is false negative, given the customer only has an issue with one sample there is no indication lot 106 is failing to meet the specificity claim in the ifu.The clinical sensitivity and specificity section of the advia centaur xp anti-hbs2 instructions for use (ifu) (10629819 revision k) lists the 95% confidence interval (ci) for resolved relative specificity as 98.34% - 99.88%, so a certain number of false positives can be expected with a given lot.If the advia centaur xp ahbs2 lot 104 result is false positive, given the customer only has an issue with one sample there is no indication lot 104 is failing to meet the specificity claim in the ifu.The cause of the discrepant result when using advia centaur xp ahbs2 lots 104 and 106 could not be determined but hsc cannot rule out a sample issue or normal assay performance.Based on the investigation, no product problem was identified.The instrument is performing within specifications.No further evaluation of the device is required.Mdr 1219913-2019-00100 supplemental report 2 (initial result), mdr 1219913-2019-00101 supplemental report 2 (repeat 1 result), and mdr 1219913-2019-00102 supplemental report 2 (repeat 2 result) were filed for the same event.
|