| Catalog Number |
515302 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/14/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Medical device expiration date: unknown.(b)(6).Device evaluated by mfr: a device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.Device manufacture date: unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that bd phaseal¿ secondary set c61 was leaking.The following information was provided by the initial reporter: leaking c62 - more information required.
|
| |
|
Event or Problem Description
|
|
It was reported that bd phaseal¿ secondary set c61 was leaking.The following information was provided by the initial reporter: leaking c62 - more information required.
|
| |
|
Additional Manufacturer Narrative
|
|
Dhr review was done and no issues were reported during production of this lot.No sample received for this complaint, picture was not provided.Ct scan was done on leaking sample which was segregated during production.After the scan additional tests were done on spike component and concentricity of lower part of spike component caused molding deficit on one side of component.Capa#891423 was initiated.
|
| |
|