• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEODENT - JJGC S.A. TITAMAX IMPLANT WITH MOUNT 3.3X13 MM; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEODENT - JJGC S.A. TITAMAX IMPLANT WITH MOUNT 3.3X13 MM; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Catalog Number 109.142
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Failure of Implant (1924); Tissue Damage (2104)
Date of Event 04/10/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.The investigation of the complaint was carried out considering the analysis of the information given by the dentist in the guarantee form, visual conference of the product returned by the dentist and the evaluation of relevant production records.
 
Event or Problem Description
(b)(4) - the dentist reported that, 9 years after the dental implant was installed in intraoral region #24, its fracture was observed.The patient presented bone type i.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITAMAX IMPLANT WITH MOUNT 3.3X13 MM
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
NEODENT - JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
MDR Report Key8658713
Report Number3008261720-2019-02387
Device Sequence Number14022175
Product Code DZE
UDI-Device Identifier07898237566672
UDI-Public07898237566672
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K101207
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2019
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 05/31/2019
Report Date (Section F) 05/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/03/2014
Device Catalogue Number109.142
Device Lot Number800005165
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2019
Was the Report Sent to FDA? No
Distributor Facility Aware Date05/31/2019
Event Location Other
Date Report to Manufacturer05/09/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/09/2019
Initial Report FDA Received Date05/31/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR
-
-