| Model Number |
DTMC1D1 |
| Medical Device Problem Codes |
Inappropriate/Inadequate Shock/Stimulation (1574); Device Sensing Problem (2917)
|
| Health Effect - Clinical Codes |
Syncope (1610); Shock from Patient Lead(s) (3162)
|
| Date of Event |
05/09/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient may have been inappropriately shocked by the cardiac resynchronization therapy defibrillator (crt-d) for an atrial arrhythmia with concurrent fast ventricular tachycardia.Right ventricular (rv) lead oversensing and short ventricular intervals were also noted upon device interrogation.The patient had reportedly been syncopal during the time of these events.The crt-d and rv lead remain in use.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Follow-up concluded that the patient was appropriately shocked for a double tachycardia per the electrophysiologist.No action was taken.
|
| |
|