• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. RUBY COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. RUBY COIL; HCG, KRD Back to Search Results
Catalog Number RBY2C2060
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/02/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2019-01070.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the right hypogastric artery using ruby coils and a lantern delivery microcatheter (lantern).During the procedure, the physician advanced a ruby coil to the distal end of the lantern and experienced resistance.The ruby coil was then retracted and re-advanced within the lantern, however, resistance was experienced when retracting and re-advancing near the distal end of the lantern.Subsequently, the ruby coil became stuck inside the lantern.An attempt was made to remove the ruby coil from the lantern, but the ruby coil was starting to become damaged.Therefore, the physician decided to remove the ruby coil and lantern together from the patient.Upon removal, it was noted that the lantern was stretched and possibly kinked at the distal end.The ruby coil was then removed from the lantern and was found to be stretched.Both devices were no longer used in the procedure.The procedure was completed using another lantern and additional ruby coils.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key8659134
Report Number3005168196-2019-01069
Device Sequence Number2275306
Product Code HCG
UDI-Device Identifier00814548019284
UDI-Public00814548019284
Combination Product (Y/N)Y
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K173614
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/02/2019
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/06/2026
Device Catalogue NumberRBY2C2060
Device Lot NumberF83909
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/02/2019
Initial Report FDA Received Date05/31/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age76 YR
Date Report Sent to FDA01/01/2005
-
-