Model Number MSPL0507
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Date of Event 05/02/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr is being submitted due to the reported suture pull through that occurred during the initial placement of the acell device.The device was not used, the surgeon decided to close the hernia primarily which resulted in a delay (increased surgery time).The patient did not experience any complication due to the suture pull through.The device was returned to acell for further investigation.A review of the manufacturing records for this lot identified no deviations in the production process that fall outside of specification.The device was manufactured and distributed sterile in compliance with acell's operating procedures and federal, state, and local laws and regulations.
 
Event or Problem Description
On (b)(4) 2019, acell, inc.Became aware that suture pull through occurred with an acell device during the reinforcement of a umbilical hernia repair.The device was not used.The surgeon decided to close the hernia primarily which resulted in a surgical delay (increased surgery time).The patient did not experience any complication due to the suture pull through nor the increased surgery time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GENTRIX SURGICAL MATRIX PLUS
Common Device Name
GENTRIX SURGICAL MATRIX PLUS
Manufacturer (Section D)
ACELL, INC.
6640 eli whitney dr.
columbia MD 21046
MDR Report Key8659456
Report Number3005920706-2019-00010
Device Sequence Number2286708
Product Code FTM
UDI-Device Identifier00386190001226
UDI-Public(01)00386190001226(17)200731(10)013335
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K182259
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2020
Device Model NumberMSPL0507
Device Catalogue NumberMSPL0507
Device Lot Number013335
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/06/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/02/2019
Initial Report FDA Received Date05/31/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-