• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.8MM GUIDE WIRE WITH FLUTES 450MM; PIN,FIXATION,SMOOTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.8MM GUIDE WIRE WITH FLUTES 450MM; PIN,FIXATION,SMOOTH Back to Search Results
Model Number 02.207.001
Medical Device Problem Code Device Damaged Prior to Use (2284)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The 2.8 mm guide wire with flutes 450 mm was received with the distal tip of the guide wire bent.This is consistent with the reported complaint condition, thus confirming the complaint.The guide wire was still within the packaging and was received bent.Conclusive evidence of the guide wire bending has been reached therefore a dimensional analysis will not be performed.There is no indication that a design or manufacturing issue contributed to the complaint.While no definitive root cause could be determined it is possible that the device encountered unintended forces during packaging transit.No new malfunctions were observed during the course of this investigation.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device history: part number: 2.207.001.Lot number: h793553.Part mfg date: 20-dec-2018.Part exp.Date: n/a.Manufacturing location: (b)(4).A review of the device history record revealed no complaint related anomalies.The device history record shows this lot of 2.8 mm guide wire with flutes 450 mm was processed through the normal manufacturing and inspection operations with no non-conformance noted.This lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.The review of the raw material device history record revealed this lot met all specifications with no non-conformance noted.This raw material lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, the guide wire with flutes was bent.There was no patient involvement.This report is for one (1) 2.8 mm guide wire with flutes 450 mm.This is report 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.8MM GUIDE WIRE WITH FLUTES 450MM
Common Device Name
PIN,FIXATION,SMOOTH
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key8660228
Report Number2939274-2019-58405
Device Sequence Number2274708
Product Code HTY
UDI-Device Identifier10886982051721
UDI-Public(01)10886982051721
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
PRE-AMEND
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number02.207.001
Device Catalogue Number02.207.001
Device Lot NumberH793553
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2019
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/20/2019
Initial Report FDA Received Date05/31/2019
Patient Sequence Number1
-
-