| Model Number |
2258 |
| Medical Device Problem Code |
Low Readings (2460)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been returned to the local facility, but has not yet been received at the main office for evaluation.Once the device has been returned and investigated, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported the sensor sometimes gives normal readings but mostly gives low readings.No patient impact or consequences were reported.
|
| |
|
Event or Problem Description
|
|
The customer reported the sensor sometimes gives normal readings but mostly gives low readings.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: the returned sensor was evaluated.During evaluation the sensor passed all visual and functional testing. during simulation testing, the sensor passed manual and preset conditions and provided accurate measurements.The sensor was determined to be functioning as designed. .
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: b3 date of event: corrected from (b)(6) 2019 to (b)(6) 2019.
|
| |
|
Event or Problem Description
|
|
The customer reported the sensor sometimes gives normal readings but mostly gives low readings.No patient impact or consequences were reported.
|
| |
|