Model Number N/A
Device Problem Power Problem (3010)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/09/2019
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was not required to be reviewed as per company standard operating procedure since the device manufacture date is greater than one year from the event date.A getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit and observed the locking mechanism was stuck causing the iabp unit to stay in rescue mode.The fse removed, cleaned and reattached the handle on the underside of the cart console to correct the issue.Subsequently, the fse completed all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.The full name of the initial reporter has been abbreviated due to field character limit; the full name should read (b)(6).
 
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) was running on the batteries while the iabp unit was connected to a/c power.It was later reported that the iabp unit generated a low battery alarm and the display showed the rescue icon there was no patient harm and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE E/F PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key8660466
MDR Text Key148325422
Report Number2249723-2019-00893
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 05/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number0998-00-0800-55
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/10/2019
Initial Date FDA Received05/31/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-