| Catalog Number |
SCLP-320 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
05/02/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformance noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.The reason for reoperation: capsular contracture, baker grade iii.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side capsular contracture, baker grade iii.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: fold creases, deformation, cloudy gel and bubbles in gel were observed after autoclave cycle.A weight test of the explanted material was verified and it was within specification.Based on the device analysis the final assessment is: no observations found related with the complaint reported by the physician.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side capsular contracture, baker grade iii.Device has been explanted.
|
| |
|