| Model Number |
38270 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
05/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a foreign material inside the catheter was observed.An imager angiographic catheter was selected for use for a carotid angiogram in the carotid artery.During preparation when attempting to flush the catheter, it was noted that there was a piece of plastic inside the catheter that popped out upon flushing.The same issue with two other imager catheters.The devices did not enter the patients body.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Mfg.Site email - (b)(4).Device evaluated by mfr: the imager diagnostic catheter was returned for evaluation.The shaft and the remainder of the device were checked for damage.Visual examination showed no damage to the shaft.The device was returned with a piece of plastic, that was reported of being extruded from the shaft during the flushing process.The complaint was confirmed for fm in the lumen.Inspection of the remainder of the device, revealed no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that a foreign material inside the catheter was observed.An imager angiographic catheter was selected for use for a carotid angiogram in the carotid artery.During preparation when attempting to flush the catheter, it was noted that there was a piece of plastic inside the catheter that popped out upon flushing.The same issue with two other imager catheters.The devices did not enter the patients body.There were no patient complications reported.
|
| |
|