| Catalog Number |
PLE60A |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch: unknown.A manufacturing record evaluation was performed for the finished device and no non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date, a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that during a laparoscopic gastric sleeve procedure, when doctor use the device, when firing the device blade cannot be retracted to its original position, so use manual retraction.Unknown if the procedure was delayed.Unknown if the procedure was completed successfully.There were no adverse consequences for the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Batch # r57887.Investigation summary: the analysis found that one ple60a device was returned with no apparent damage and with no cartridge reload present.The manual override door was noted to be out of position; the override lever was up which denotes that the knife was manually returned to home position.It should be noted after the manual override system is used the instrument is disabled and cannot be used for any subsequent firings.The bailout system was reset and then, it was tested for functionality in the straight position with a test cartridge reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples were noted to have the proper b-form shape.Event could not be confirmed as the device opened and closed without any difficulties noted.Once the device completed the firing sequence the knife returned home as intended.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|