Device Displays Incorrect Message (2591); Fail-Safe Problem (2936)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019, the reporter contacted animas, alleging a call service alarm (call service alarm issue) issue; cs 069 alarm.There is no indication the alleged product issue caused or contributed to an adverse event.This complaint is being reported because the issue may result in a long-term cessation of insulin delivery if the user is unable to resolve the alarm.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the pump has been returned and evaluated by product analysis on 01-aug-2019 with the following findings: a review of the black box data and alarm history revealed multiple call service alarms had occurred.During investigation, the pump alarmed with a call service (cs 069) at power up.The call service 69 failure was written to the black box as a call service 87 failure.A language corruption occurred at a component on the printed circuit board resulting in a call service alarm.Unrelated to the original complaint, investigation revealed a crack in the battery compartment.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
|
| |
|