ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Codes Device Displays Incorrect Message (2591); Fail-Safe Problem (2936)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event or Problem Description
On (b)(6) 2019, the reporter contacted animas, alleging a call service alarm (call service alarm issue) issue; cs 069 alarm.There is no indication the alleged product issue caused or contributed to an adverse event.This complaint is being reported because the issue may result in a long-term cessation of insulin delivery if the user is unable to resolve the alarm.
 
Additional Manufacturer Narrative
Device evaluation: the pump has been returned and evaluated by product analysis on 01-aug-2019 with the following findings: a review of the black box data and alarm history revealed multiple call service alarms had occurred.During investigation, the pump alarmed with a call service (cs 069) at power up.The call service 69 failure was written to the black box as a call service 87 failure.A language corruption occurred at a component on the printed circuit board resulting in a call service alarm.Unrelated to the original complaint, investigation revealed a crack in the battery compartment.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
MDR Report Key8663494
Report Number2531779-2019-03441
Device Sequence Number19378650
Product Code OYC
UDI-Device Identifier10840406100051
UDI-Public10840406100051
Combination Product (Y/N)Y
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/25/2019
Was the Report Sent to FDA? Yes
Device Age42 MO
Initial Date Received by Manufacturer 05/31/2019
Supplement Date Received by Manufacturer05/31/2019
Initial Report FDA Received Date06/03/2019
Supplement Report FDA Received Date08/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age49 YR
Patient Weight204
-
-