| Model Number |
7887-050 |
| Medical Device Problem Codes |
Electrical Shorting (2926); Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/30/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the levo arm was in use and locked while a patients head was connected to the skull clamp.The levo arm released and the patients head dropped without any release buttons being depressed.
|
| |
|
Additional Manufacturer Narrative
|
|
The cause of the unintended motion was a short circuit of the control circuit due to fluid ingress.Fluid ingress path appears to have originated from around the upper release button at the control handle.This fluid caused a temporary short circuit.During inspection the fluid had dried out and the short circuit was not measurable.
|
| |
|
Event or Problem Description
|
|
It was reported the levo arm was in use and locked while a patients head was connected to the skull clamp.The levo arm released and the patients head dropped without any release buttons being depressed.
|
| |
|
Additional Manufacturer Narrative
|
|
The cause of the unintended motion was a short circuit of the control circuit due to fluid ingress.Fluid ingress path appears to have originated from around the upper release button at the control handle.This fluid caused a temporary short circuit.During inspection the fluid had dried out and the short circuit was not measurable.
|
| |
|
Event or Problem Description
|
|
It was reported the levo arm was in use and locked while a patients head was connected to the skull clamp.The levo arm released and the patients head dropped without any release buttons being depressed.
|
| |
|