| Catalog Number |
702-04-54E |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
05/06/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
Rep reported "patient implanted with 10* trident constrained acetabular liner with 22mm/0+ inner head.Liner disassociated from shell upon patient transfer from surgical table to hospital bed, post anesthesia wakeup and site closure.Patient had to be reinduced and re-opened for immediate revision.Constrained liner and inner head were revised to new implants; surgeon requested that primary implants be investigated.Update: "no allegations against the head; surgeon question locking mechanism." right side.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: d10/h3 - the trident shell was not returned to stryker additional manufacturer narrative: reported event: an event regarding disassociation involving a trident shell was reported.The event was not confirmed.Method & results: -device evaluation and results: not performed as product was not returned -clinician review: a review of the provided medical records by a clinical consultant stated the following comment: need operative reports, office/clinical reports, examination of components, etc.-device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the reported lot.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, office/clinical reports, examination of components, etc.Are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.H3 other text : device not available.
|
| |
|
Event or Problem Description
|
|
Rep reported "patient implanted with 10* trident constrained acetabular liner with 22mm/0+ inner head.Liner disassociated from shell upon patient transfer from surgical table to hospital bed, post anesthesia wakeup and site closure.Patient had to be reinduced and re-opened for immediate revision.Constrained liner and inner head were revised to new implants; surgeon requested that primary implants be investigated.Update: "no allegations against the head; surgeon question locking mechanism." right side.
|
| |
|