DAVIS & GECK CARIBE LTD SOFSILK; SUTURE, NONABSORBABLE, SILK Back to Search Results
Model Number GS-30-M
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Radiation Exposure, Unintended (3164)
Date of Event 05/02/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a laparoscopic procedure, the needle, broken during detaching the suture and an x-ray was performed to confirm if all pieces were located.A new suture was used to complete the procedure.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the opened samples noted one needle.The needle was received broken at the tip and received separated from the rest of the needle.Upon microscopic inspection, instrument marks were observed near the break site.As no sealed samples were returned, a bend moment test could not be performed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.The needle should always be held in the middle where the diameter is greatest.Grasping the needle near the swaged end or the tip could cause bends or breaks, or damage the connection between the needle and suture.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFSILK
Common Device Name
SUTURE, NONABSORBABLE, SILK
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 18750
DO  18750
MDR Report Key8665386
Report Number9612501-2019-01064
Device Sequence Number2103209
Product Code GAP
UDI-Device Identifier20884521084848
UDI-Public20884521084848
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K981128
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2022
Device Model NumberGS-30-M
Device Catalogue NumberGS-30-M
Device Lot NumberD7L2094X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/19/2019
Initial Date Received by Manufacturer 05/10/2019
Supplement Date Received by Manufacturer07/26/2019
Initial Report FDA Received Date06/04/2019
Supplement Report FDA Received Date08/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-