| Model Number |
MMT-722NAL |
| Medical Device Problem Codes |
Device Difficult to Program or Calibrate (1496); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/28/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that their insulin pump had up and down arrow key not working.The customer¿s blood glucose was 284 mg/dl at the time of incident.Customer reports they getting the message of battery out limit.Customer reports they were able to clear the alarm.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with intermittent button response due to corroded keypad traces.No button error alarm during testing.Device received with all operating currents within specification and passed a21 error test and displacement test.No unexpected battery out limit alarm anomalies noted.
|
| |
|