• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, INC AFFIRM PRONE BIOPSY SYSTEM, 3D; BREAST BIOPSY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC AFFIRM PRONE BIOPSY SYSTEM, 3D; BREAST BIOPSY SYSTEM Back to Search Results
Model Number PBX-SYS-AFFIRM-3D
Medical Device Problem Code Unintended System Motion (1430)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/16/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the c-arm is moving on its own.Rebooting the system did not resolve the issue.A field engineer was dispatched to the site and cycled the lock on the c-arm.Once this was completed, the system was working as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFIRM PRONE BIOPSY SYSTEM, 3D
Common Device Name
BREAST BIOPSY SYSTEM
Manufacturer (Section D)
HOLOGIC, INC
36 & 37 apple ridge road
danbury CT 06810
MDR Report Key8667137
Report Number1220984-2019-00061
Device Sequence Number2103690
Product Code IZH
UDI-Device Identifier15420045506510
UDI-Public15420045506510
Combination Product (Y/N)N
Initial Reporter StateDC
Initial Reporter CountryUS
PMA/510(K) Number
K153486
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPBX-SYS-AFFIRM-3D
Device Catalogue NumberPBX-SYS-AFFIRM-3D
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/16/2019
Initial Report FDA Received Date06/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-