| Catalog Number |
23510050 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/15/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon completion of investigation, additional information will be submitted in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that, during a primary procedure on a left femur, the threads of the strike plate broke off in the targeted upon impaction.Procedure was completed successfully with no delay or adverse consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: section d4 (lot#).The reported event could be confirmed.The device returned broke on the transition radii of the thread undercut to the stop shoulder of the thread connection.A microscope analysis of the surface of the device shows a brittle area, which proves that the material was subjected to a significant amount of stress, which lead to it breaking without significant plastic deformation.This leads to the conclusion that this event was due to a user related root cause.The analysis of the device resulted in finding that the thread of the strike plate had been damaged by excessive load application resulting from incorrect engagement in the counterpart.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any further information is provided, the investigation report will be updated.
|
| |
|
Event or Problem Description
|
|
It was reported that, during a primary procedure on a left femur, the threads of the strike plate broke off in the targeter upon impaction.Procedure was completed successfully with no delay or adverse consequences.
|
| |
|