AGA MEDICAL CORPORATION AMPLATZER DUCT OCCLUDER; OCCLUDER, PATENT DUCTUS, ARTERIOSUS Back to Search Results
Model Number 9-PDA-006
Medical Device Problem Code Retraction Problem (1536)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/02/2019
Type of Reportable Event Serious Injury
Event or Problem Description
Related manufacturing report number: 2135147-2019-00144.On (b)(6) 2019, a 10/8mm amplatzer duct occluder (ado) was selected for implant using a 7f amplatzer torqvue delivery system.The device was attempted to be positioned twice, but the device embolized into the pulmonary artery while attached to the delivery cable.The device was attempted to be re-sheathed, but the tip of the delivery system was invaginated preventing a successful recapture.The device was surgically removed and the patent ductus arteriosus was ligated.The patient was recovery and scheduled to be discharged.
 
Additional Manufacturer Narrative
An event of embolization was reported.The investigation confirmed the device met functional and dimensional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.The cause of the reported incident could not be conclusively determined.
 
Event or Problem Description
On (b)(6) 2019, a 10/8mm amplatzer duct occluder (ado) was selected for implant using a 7f amplatzer torqvue delivery system.The device embolized into the pulmonary artery twice while attached to the delivery cable.The tip of the delivery sheath was observed to be invaginated and the physician was unable to re-sheath the device.The patient was referred to surgery and the ado was removed from the patient.On (b)(6) 2019, the patient was reported to be discharged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER DUCT OCCLUDER
Common Device Name
OCCLUDER, PATENT DUCTUS, ARTERIOSUS
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
MDR Report Key8669057
Report Number2135147-2019-00145
Device Sequence Number254500
Product Code MAE
UDI-Device Identifier00811806011134
UDI-Public00811806011134
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P020024
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/29/2020
Device Model Number9-PDA-006
Device Catalogue Number9-PDA-006
Device Lot Number5008713
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/06/2019
Supplement Date Received by Manufacturer06/25/2019
Initial Report FDA Received Date06/04/2019
Supplement Report FDA Received Date07/23/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age7 MO
-
-