| Model Number |
21102-9 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
01/03/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The attachment of the complaint states that the patient fell out of bed, which produced the plate breakage.The device history record (dhr) of this plate 21102-9 were inspected and showed no deviations.The quality forms met all dimensional and visual criteria at the time of release with no issues documented during the manufacturing that would contribute to this complaint condition.A review of the raw material device history record revealed no deviation.The material was determined to be conforming and was used as is per product development approval.
|
| |
|
Event or Problem Description
|
|
It was reported that a straight plate, 3.5mm, 9-hole was determined to be broken postoperatively.Original implant date (b)(6) 2018.A revision surgery was performed on (b)(6) 2019.
|
| |
|