| Model Number |
0170SI18 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the foley catheter leaked.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A potential root cause for this failure could be "poor release of balloon during oven drying" and potential failure mode "non-concentric balloon with no deflection".The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the foley catheter leaked.
|
| |
|
Search Alerts/Recalls
|