| Model Number |
323.46 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is a synthes employee.Device is not distributed in the united states, but is similar to device marketed in the usa.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a routine incoming inspection on (b)(6) 2019, it was observed that the universal drill guide had a stuck screw.There was no patient involvement.This report is for a universal drill guide.This is report 1 of 2 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: part: 323.460, lot: l261406, manufacturing site: hägendorf, release to warehouse date: february 09, 2017.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Service & repair evaluation: the customer reported the device had a stuck screw.The repair technician reported the drill sleeve is damage and spring is missing.Damaged component is the reason for repair.The cause of the issue is unknown.The following parts were replaced: spring, drill sleeve, and 4.5 mm universal drill guide.The item was repaired per the inspection sheet, passed synthes final inspection, and will be returned to the customer upon completion of the service and repair process.The evaluation was confirmed.The device was deemed serviceable and will be returned to the customer.No design or manufacturing issues were identified therefore it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|