• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS, INC MICRO-INTRODUCER KITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS, INC MICRO-INTRODUCER KITS Back to Search Results
Catalog Number H787065970525
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 05/17/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was reported that the disposable device was discarded by the user and is not available to be returned to the manufacturer for evaluation.An investigation into the root cause of this incident is currently in progress.The results of the investigation will be sent via a follow up medwatch.Complaint # (b)(4).
 
Event or Problem Description
Venus access on a patient using 4f stiff micro introducer kit.While they were putting the introducer on the wire, it (the introducer) appeared to have sheared off of at the tip.The treating physician was unable to remove the tip from the patient's subcutaneous tissue.It was reported that the patient suffered no serious harm or injury due to this event.It was reported the defective disposable device is not available for return to the manufacturer as it was disposed of by the user.
 
Additional Manufacturer Narrative
As the reported defective device was not returned, angiodynamics is unable to perform a device evaluation.Without a device evaluation, angiodynamics is unable to confirm the reported complaint description or to determine a root cause.Angiodynamics' supplier of this device was made aware of the incident.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.The instructions for use, which is supplied to the end user with this catalog number, lists perforation of a vessel or viscus, laceration of a vessel or viscus, extravasation, nerve damage, and embolism as potential complications.A review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO-INTRODUCER KITS
Common Device Name
MICRO-INTRODUCER KITS
Manufacturer (Section D)
ANGIODYNAMICS, INC
603 queensbury avenue
queensbury NY 12804
MDR Report Key8670488
Report Number1319211-2019-00048
Device Sequence Number16204412
Product Code DQT
UDI-Device IdentifierH787065970525
UDI-PublicH787065970525
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K051655
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberH787065970525
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/20/2019
Supplement Date Received by Manufacturer05/20/2019
Initial Report FDA Received Date06/05/2019
Supplement Report FDA Received Date09/27/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age83 YR
-
-