| Catalog Number |
COVSCD396 |
| Medical Device Problem Codes |
Smoking (1585); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/09/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not yet been returned for analysis.Manufacturer's ref.No: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure with a sonicision cordless ultrasonic dissector and the tip of device began smoking and a piece of the tip detached during procedure.The white coating on the tip of the device started to come apart during use and possibly melted off.Tiny fragment of white plastic coating completely detached inside of the patient.Tip of instrument looked charred.All visible material was retrieved successfully.There was no patient consequence.This event has been assessed as reportable.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received via medsun/user facility report (b)(4) on (b)(6) 2019.It was noted that a laparoscopic surgery was performed on a (b)(6)-year-old female patient, weighing (b)(6) kgs, not hispanic/latino and race of white.Updated sections: patient identifier, sex, weight, ethnicity, describe event or problem., initial reporter also sent report to fda, report source, date received by manufacturer.Manufacturer's ref.No: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that a patient underwent an unknown procedure with a sonicision cordless ultrasonic dissector and the tip of device began smoking and a piece of the tip detached during procedure.The reported issue could not be confirmed because the product said to be involved was not received for evaluation.As the device was not returned for evaluation the visual and functional inspections could not be performed.Per the failure modes and effects analysis (fmea) for this device, possible cause for this hazard is operator error - activation of the device without tissue between the blade and the pad.A manufacturing record evaluation was performed for the finished device lot# 2081413, and no non-conformances were identified.Manufacturer's ref.No: (b)(4).
|
| |
|