• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752439
Device Problems Break (1069); Detachment of Device or Device Component (2907)
Patient Problem Tissue Damage (2104)
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A physician reported that the tip of the vitrector came off inside the patient's eye.The sclerotomy site was enlarged to remove the broken off tip with a forcep.The procedure was completed.
 
Manufacturer Narrative
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.A complaint history examination indicates there are no additional complaints associated with the lot for the reported issue.No sample has been returned for evaluation for the report of tip came off inside eye; therefore, the condition of the product could not be verified.A sample was not returned and the device history record review of the lot number provided indicated the product was processed and released according to the product¿s acceptable criteria, therefore, the root cause for the customer complaint issue cannot be determined.The exact root cause for this complaint is unknown as no sample was returned; therefore, specific action with regards to this complaint cannot be taken.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Any non-conformances found are removed from the lot and scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
MDR Report Key8674373
MDR Text Key147449669
Report Number1644019-2019-00087
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Remedial Action Modification/Adjustment
Type of Report Initial,Followup
Report Date 07/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/06/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2021
Device Catalogue Number8065752439
Device Lot Number2256157H
Was Device Available for Evaluation? No
Date Manufacturer Received06/14/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-