|
(b)(4).Date sent to the fda: 07/23/2019.Device evaluation summary: it was received for analysis a needle suture piece of product code z317.During visual inspection of the needle-suture piece, the swage and attachment area were noted to be as expected, and the end of suture was observed cut appears to be by use of a surgical instrument.In addition, a functional test was performed on the sample and the tensile force met the requirements.The manufacturing records were reviewed, and the manufacturing/ packaging criteria were met prior to the release of this batch.Due to the sample condition, the assignable cause of performance breakage suture, was suggests an improper handling of the sample.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of the surgical instruments, such as needle holders and forceps.
|