• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SUSTAINABILITY SOLUTIONS PHOENIX; CATHETER, RECORDING, ELECTRODE, REPROCESSED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SUSTAINABILITY SOLUTIONS PHOENIX; CATHETER, RECORDING, ELECTRODE, REPROCESSED Back to Search Results
Model Number 81105
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/12/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Upon visual inspection of the received complaint device, a bend 5 cm from the distal tip was identified.The device did not meet the curve specification.The device did not meet the planarity specification.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.The most likely root cause is mishandling subsequent to distribution, including shipping/storage conditions or improper manipulation of the catheter.The instructions for use (ifu) state: inspect the catheter for overall condition and physical integrity.Do not use the catheter if electrodes appear loose or if any damage is noted.If such problems exist, return the catheter and packaging to stryker sustainability solutions.Connect the catheter to the corresponding cable connector.Connect the cable to the correct electronic equipment for recording and/or sensing.Observe polarity of proximally located connector pins of the interface cable when connecting to the electronic equipment.Isolate any unused connector pins to reduce development of accidental current pathways to the heart.Follow a suitable electrophysiology study protocol.Do not autoclave catheter.Do not use for electrical ablation.Avoid excessive contact of handpiece with fluids, as this could adversely affect the electrical performance of the catheter.Diagnostic ep catheters are not recommended for long term pacing.The reported event will continue to be monitored through post-market surveillance.
 
Event or Problem Description
It was reported that the electrophysiology catheter was defective.The complainant is not aware of any patient injury, medical intervention, or extended procedure time.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
CATHETER, RECORDING, ELECTRODE, REPROCESSED
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS PHOENIX
10232 s. 51st st.
phoenix AZ 85044
MDR Report Key8675335
Report Number0002090040-2019-00017
Device Sequence Number2511332
Product Code NLH
UDI-Device Identifier00885825009564
UDI-Public00885825009564
Combination Product (Y/N)N
Initial Reporter StateDC
Initial Reporter CountryUS
PMA/510(K) Number
K112232
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/23/2020
Device Model Number81105
Device Catalogue Number81105RR
Device Lot Number3664317
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2019
Initial Report FDA Received Date06/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
-
-