| Model Number |
81105 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/12/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to stryker sustainability solutions for evaluation.Upon visual inspection of the received complaint device, a bend 5 cm from the distal tip was identified.The device did not meet the curve specification.The device did not meet the planarity specification.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.The most likely root cause is mishandling subsequent to distribution, including shipping/storage conditions or improper manipulation of the catheter.The instructions for use (ifu) state: inspect the catheter for overall condition and physical integrity.Do not use the catheter if electrodes appear loose or if any damage is noted.If such problems exist, return the catheter and packaging to stryker sustainability solutions.Connect the catheter to the corresponding cable connector.Connect the cable to the correct electronic equipment for recording and/or sensing.Observe polarity of proximally located connector pins of the interface cable when connecting to the electronic equipment.Isolate any unused connector pins to reduce development of accidental current pathways to the heart.Follow a suitable electrophysiology study protocol.Do not autoclave catheter.Do not use for electrical ablation.Avoid excessive contact of handpiece with fluids, as this could adversely affect the electrical performance of the catheter.Diagnostic ep catheters are not recommended for long term pacing.The reported event will continue to be monitored through post-market surveillance.
|
| |
|
Event or Problem Description
|
|
It was reported that the electrophysiology catheter was defective.The complainant is not aware of any patient injury, medical intervention, or extended procedure time.These are commonly used devices that are readily available.
|
| |
|
Search Alerts/Recalls
|
|
|