Catalog Number 388.394
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/09/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2019, a cervical posterior fixation surgery at c4 was performed.During the surgery, when the surgeon tried to drill for the pilot hole for an unknown lateral mass screw by sizing up to 16 or 18mm, the drill bit broke off.Then, the surgeon used an alternative drill bit to continue drilling, however, the drill bit was also broken.Thus, the surgery was completed by using another alternative drill bit.All broken fragments were removed.There was a 5-minute surgical delay and there was no adverse consequence to the patient.Concomitant devices reported: unknown handle (part # unknown, lot # unknown, quantity of 1).This is report 2 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: a review of the device history record device history lot part number: 388.394, synthes lot number: u200415, supplier lot number: n/a, release to warehouse date: 05 jun 2014, expiration date: n/a, supplier: (b)(4).No ncrs were generated during production.Device history batch null, device history review review of the device history record showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.Investigation summary investigation selection investigation site: cq zuchwil selected flow: damaged - broken.Visual inspection: the received drill bit shows that that approx.14mm from the fluted tip section is broken off.The cutting edges at the tip are slightly worn.The shaft and coupling are otherwise still in good condition.Dimensional inspection: conclusion: the measuring result does show conformity.Drawing/specification review: the manufacturing review shows that the production procedures were according to the specifications and there were no issues that would contribute to this complaint condition.Failure in material or production could not be detected.Summary: the received condition of the device is concordant with the complaint description and the complaint condition is confirmed.Based on the provided information we are not able to determine the exact cause of this complaint.We only can assume that a mechanical overloading situation, for example metallic contact or lateral stress, has caused the breakage.Please also note; blunt drill bits require more mechanical power during the application, therefore we would like to draw your attention on page 4 in the leaflet ¿important information¿ check instruments for sound surfaces, and correct adjustment and function.Do not use severely damaged instruments, instruments with unrecognizable markings, corrosion, or blunt cutting surfaces.We conclude that the cause of failure is not due to any manufacturing non-conformances.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed as no product related issue could be detected.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.4MM DRILL BIT/QC
Common Device Name
BIT,DRILL
Manufacturer (Section D)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key8675981
Report Number8030965-2019-65012
Device Sequence Number2130006
Product Code HTW
UDI-Device Identifier07611819752931
UDI-Public(01)07611819752931
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number388.394
Device Lot NumberU200415
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2019
Initial Date Received by Manufacturer 05/09/2019
Supplement Date Received by Manufacturer06/12/2019
Initial Report FDA Received Date06/06/2019
Supplement Report FDA Received Date07/08/2019
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - HANDLES
-
-