| Catalog Number |
168-420 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
05/08/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The reason for reoperation: deflation and "pain".Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a left side deflation and "pain".Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: creases, white particles and one opening.A leak test was performed which identified an opening.A microscopic analysis was performed which identified one striated opening.A fill inspection was performed and identified no blockage in the valve.Based on the device analysis the final assessment is: one striated opening assessed as surgical damage.
|
| |
|