| Catalog Number |
32-118 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The date of the incident is unknown.The return of the device is anticipated.Once we have received the device and evaluated.A follow-up report will be submitted.The delay in the fda receiving this complaint was due to a change in esignature certificates.
|
| |
|
Event or Problem Description
|
|
When the surgeon cut the pin, the pin cutter blade broke into multiple pieces and went into the patient's right ankle.The surgeon had to do an ankle scope on the patient to retrieve the foreign bodies.Using x-ray and the scope the all of the pieces were retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
After multiple attempts, the device was not returned for evaluation.Based on the lot number provided in the intial report of the complaint, this device was purchased in dec 2016 and had been in use for approximately 2 years at the time that the complaint was reported.Because the device not been returned, a root cause could not be determined, therefore, this can be seen as the final report.If additional information is obtained that alleges any additinal patient involvement or the need for corrective actions a follow-up report will be submitted.
|
| |
|