| Model Number |
ZXT150 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: exact date is unknown, not provided, but best estimate date is between (b)(6) 2018 - (b)(6) 2019.(b)(4).Device evaluation: product testing could not be performed because the product was not returned.The reported complaint was not confirmed.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.The search revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) were reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that tecnis symfony toric intraocular lens (model zxt150 +22.0 diopter) was explanted from patient¿s left eye in secondary surgical procedure because of the patient being unhappy and experiencing glare.A model zct150 +22.0 diopter lens was used as the replacement lens for the patient.Additional information was received and it was learnt that incision was enlarged during the procedure and the explanted lens was discarded.Patient was doing fine at discharge.Visual acuity preop 20/20.1st procedure visuals acuity preop 20/20.2nd procedure (exchange)visual acuity preop 20/30 one day post op.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|
Search Alerts/Recalls
|