Catalog Number X0212513
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/13/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient underwent scoliosis correction surgery at t3-l3.The patient had a stiff double major scoliotic curve.Intra-op, when surgeon applied force to the segmental derotators, the screw head got deformed in the coronal plane; which ultimately plowed through the lateral vertebral wall of left side l2 (away from the spinal cord).The screw was then completely explanted.This level was left un-instrumented after the event.No patient complications were reported.
 
Additional Manufacturer Narrative
Product analysis: visual and optical inspection did not reveal any damage to the head or threads of the bone screw.Dimensional check confirmed the bone screw was made to print.It was not possible to determine root cause of the event.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key8678017
Report Number1030489-2019-00611
Device Sequence Number19316703
Product Code KWP
UDI-Device Identifier00643169606951
UDI-Public00643169606951
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K130646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberX0212513
Device Lot Number0677130W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2019
Initial Date Received by Manufacturer 05/13/2019
Supplement Date Received by Manufacturer07/01/2019
Initial Report FDA Received Date06/07/2019
Supplement Report FDA Received Date07/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age11 YR
Patient Weight44
-
-