• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP CONSOLE; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP CONSOLE; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE Back to Search Results
Model Number 60-03-75
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/06/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient information was not provided.Livanova (b)(4) manufactures the centrifugal pump console.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue as the unit was not turned off since the malfunction occurred.The unit was then switched off and the panel was opened to check the cable connections.The device rebooted and the reported issue could not be reproduced anymore.Therefore livanova (b)(4) requested the device back for further investigation.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event or Problem Description
Livanova (b)(4) received a report that the display of a centrifugal pump console had a blue screen during an ecmo procedure.There was no report of patient injury.
 
Additional Manufacturer Narrative
H.10: the reported issue could be reproduced during the investigation at the manufacturer site.The root cause for the blue screen is a defective contrast potentiometer.
 
Event or Problem Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP CONSOLE
Common Device Name
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8678256
Report Number9611109-2019-00434
Device Sequence Number2472409
Product Code KFM
UDI-Device Identifier04033817901020
UDI-Public(01)04033817901020(11)180320
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K020571
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 10/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number60-03-75
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/10/2019
Supplement Date Received by Manufacturer09/04/2019
Initial Report FDA Received Date06/07/2019
Supplement Report FDA Received Date10/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-