Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Cardiac Perforation (2513)
Event Type  Injury  
Manufacturer Narrative
Concomitant devices: flexability¿ ablation catheter, swartz¿ introducer, agilis¿ introducer.The results of the investigation are inconclusive since the device was not returned for analysis.Additionally, no lot number was provided so a review of the device history record (dhr) was not possible.Based on the information received, the cause of the reported pericardial effusion could not be conclusively determined.Per the ifu, cardiac perforation is a known risk during the use of this device.
 
Event Description
Related manufacturing ref: 3005334138-2019-00297, 3005334138-2019-00298, 182269-2019-00078.The following was published in the journal of arrhythmia in an article titled, "pulmonary vein isolation plus left atrial posterior wall isolation and additional nonpulmonary vein trigger ablation using high-dose isoproterenol for long-standing persistent atrial fibrillation" by tomomasa takamiya, md, junichi nitta, md, akira sato, md, et al., on feb.18, 2019."the median procedure time was 183 (q1-q3; 155-220) minutes.Complications occurred in 7 of 226 (3%) procedures; 4 (2%) patients had pericardial effusion that required percutaneous drainage, 2 (1%) had gastroparesis, and 1 (0.4%) patient had an esophageal ulcer.All patients were conservatively treated without long-term sequelae.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOL PATH¿ IRRIGATED ABLATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
2375 morse ave
irvine CA 92614
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
2375 morse ave
irvine CA 92614
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key8678538
MDR Text Key147323017
Report Number2030404-2019-00041
Device Sequence Number1
Product Code OAD
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P110016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 06/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/14/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-