ENCORE MEDICAL L.P. ENCORE HUMERAL SHOULDER STEM; ENCORE REVERSE SHOULDER HUMERAL STEM, STD, SZ12X108MM Back to Search Results
Model Number 530-12-108
Health Effect - Clinical Code No Code Available (3191)
Date of Event 05/08/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was humeral fracture.The previous surgery and the surgery detailed in this investigation occurred 14 days apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.Initial or prolonged hospitalization was required.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to humeral fracture.There were no findings during this investigation that indicate that the reported device was defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.
 
Event or Problem Description
Revision surgery - due to a humeral fracture distal to the humeral stem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCORE HUMERAL SHOULDER STEM
Common Device Name
ENCORE REVERSE SHOULDER HUMERAL STEM, STD, SZ12X108MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key8678981
Report Number1644408-2019-00501
Device Sequence Number2114965
Product Code PHX
UDI-Device Identifier00888912168427
UDI-Public(01)00888912168427
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K141990
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/25/2025
Device Model Number530-12-108
Device Catalogue Number530-12-108
Device Lot Number412T1366
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/08/2019
Initial Report FDA Received Date06/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
509-00-036, LOT 378P1194
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age70 YR
-
-