The following were reviewed as part of this investigation: patient severity, complaint and lot history review, applicable previous investigation(s), sample (if available), labeling, applicable manufacture records, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a leak was confirmed and the cause appears to be use related.The product returned for evaluation was one 5 fr dual lumen groshong nxt catheter.The damage observed in the returned sample was characteristic of over-pressurization (burst) damage.This can occur through the use of syringes smaller than 10ml, by flushing against an occlusion, or due to excessive force applied during infusion procedures.The returned product sample was evaluated and a split was observed just distal to the bifurcation.This catheter damage was typical of a burst, and the characteristics observed which supported this type of failure included: 'c' shaped split.Tensile weakness at the fracture site (due to material ballooning prior to burst).Granular fracture surface texture (typical of tearing failure modes).The nursing procedure manual can be found online at www.Bardaccess.Com and provides the following guidance: ¿do not use a syringe smaller than 10ml to flush or confirm patency.Patency should be assessed with a 10ml or larger syringe with sterile saline.Upon confirmation of patency, administration of medication should be given in a syringe appropriately sized for the does.Do not infuse against resistance.Infusion pressures should never exceed 25 psi.Smaller syringes generate more pressure than larger syringes.¿ the nursing procedure manual also provides guidance for re-establishing catheter patency in the event the catheter becomes occluded or resistance is felt during flushing.
|