| Catalog Number |
TRF-295 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
03/26/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformance's noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The reason for reoperation is capsular contracture, baker grade iii.This is a known potential adverse event addressed in the product labeling.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a left side capsular contracture baker grade iii.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified a deformation, fold crease and weight of the device to the specification.Clouds and voids were observed after autoclave disinfection process.Based on the device analysis, the final assessment is no issues found related with the manufacturing process.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a left side capsular contracture baker grade iii.Device has been explanted.
|
| |
|