| Model Number |
015-044 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/31/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent cranioplasty on (b)(6) 2019 due to a tumor.During the operation, the device's hole fractured.It was stated that it was replaced with another device on site.It extended the procedure by 10 minutes.The patient's status at the time of the report was alive-no injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned timesh plat was bent in half.The tab was not returned with the device.One of the eyelets was distorted and broken.It is unknown how or when this damage occurred.The instructions for use cautions, ¿excess force, over-bending, notching, or scratching during mesh forming could result in a mesh fracture.¿ if information is provided in the future, a supplemental report will be issued.
|
| |
|