ATRIUM MEDICAL CORPORATION OCEAN DRAIN; BOTTLE, COLLECTION, VACUUM Back to Search Results
Model Number 2002-300
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
We are in the process of performing the investigation and will submit the follow-up report once the evaluation is completed.
 
Event or Problem Description
Report received that a suction tube was found missing when user opened the package.
 
Additional Manufacturer Narrative
The ocean chest drain was returned from the field and evaluated.Upon initial inspection the suction line assembly was not present with the drain confirming the complaint.Only the patient line tube set was present.A review of the device history records show that all parts were supplied and consumed for this production build of chest drains.The build quantity was 3,510 units.The device history records also show that the quality inspection confirmed the presence of the suction line on 50 individual units picked from the production line prior to packaging.There were no non- conformances noted during the build of this ocean drain lot.Based on the results of the investigation atrium medical cannot determine that the suction line was not present when the product was shipped.Clinical evaluation: the ocean chest drainage system is indicated for the evacuation of air and/or fluid from the chest cavity or mediastinum and to help re-establish lung expansion and restore breathing dynamics.It also facilitates postoperative collection and reinfusion of autologous blood from the patient¿s pleural cavity or mediastinal area.The instructions for use advise: "do not use if the package is opened or damaged".The device is supplied sterile and the packaging should be inspected prior to use to ensure it is intact and not damaged.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCEAN DRAIN
Common Device Name
BOTTLE, COLLECTION, VACUUM
Manufacturer (Section D)
ATRIUM MEDICAL CORPORATION
40 continental blvd
merrimack NH 03054
MDR Report Key8681121
Report Number3011175548-2019-00639
Device Sequence Number19406301
Product Code KDQ
UDI-Device Identifier00650862100105
UDI-Public00650862100105
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K043582
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/04/2021
Device Model Number2002-300
Device Catalogue Number2002-300
Device Lot Number434464
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/19/2019
Initial Date Received by Manufacturer 05/17/2019
Supplement Date Received by Manufacturer07/01/2019
Initial Report FDA Received Date06/07/2019
Supplement Report FDA Received Date07/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-